Model Number MI1200 SYNCHRONY PIN
Medical Device Problem Code Impedance Problem (2950)
Health Effect - Clinical Codes Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
The user reported no hearing sensation with the device.Re-implantation is being considered but will not be performed earlier than in two to three months.
 
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event or Problem Description
The user reported no hearing sensation with the device.Re-implantation is being considered but will not be performed earlier than in two to three months.
 
Additional Manufacturer Narrative
Additional information: based on the received information, a damage to the active electrode due to excessive mechanical stress appears very likely.However, to determine an exact root cause device investigation would be necessary.Re-implantation is considered but no date has been scheduled yet.
 
Additional Manufacturer Narrative
Conclusion: based on the received information, a damage to the active electrode due to excessive mechanical stress appears very likely.However, to determine an exact root cause device investigation would be necessary.The concerned device was explanted but has not been received for investigation yet.
 
Event or Problem Description
The user reported no hearing sensation with the device.Re-implantation is being considered.
 
Additional Manufacturer Narrative
Conclusion: overload fractures in the active electrode which are consistent with an external mechanical impact were determined to be the root cause of device failure.The problems given in the recipient report appear to match the damage found.This is a final report.
 
Event or Problem Description
The user reported no hearing sensation with the device.User was reimplanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHRONY
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key21374484
Report Number9710014-2025-00134
Device Sequence Number3826015
Product Code MCM
UDI-Device Identifier09008737093969
UDI-Public(01)09008737093969
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 04/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMI1200 SYNCHRONY PIN
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/10/2025
Supplement Date Received by Manufacturer02/10/2025
02/10/2025
02/10/2025
Initial Report FDA Received Date02/13/2025
Supplement Report FDA Received Date02/19/2025
02/05/2026
04/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age4 YR
Patient SexMale
-
-