| Catalog Number |
PPH03 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
01/06/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Per user facility medwatch form, #mw5164942, the surgeon stapled hemorroidopexy with ethicon hemorrhoid pph stapler cpt 46947 and returned to the ed the following day with rectal bleeding.This was reported from home and observed in the ed.Patient to the or for anorectal exam and, although no active bleeding was noted at that time, the staple line was reinforced with sutures and she was discharged.Patient pre-op hemoglobin was 16, on arrival to emergency room on 1-7 was 14 and then to 12.4.Device is available for return.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent 2/13/2025.D4 batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: what was the grade of hemorrhoids? were the hemorrhoids circumferential or in specific quadrants? please describe.What is the surgeon¿s experience with the pph procedure? what is the surgeon¿s experience with the pph device? where in the green gap setting scale was the indicator located prior to firing (low-b, middle-b, or high-b)? did the surgeon wait 30 seconds after closing the device and then 20 seconds after firing? was a complete washer transection confirmed? were there any issues noted with staple formation at any point throughout the care of the patient? if so, please describe the shape and location.What is the current patient status? this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) date sent: 4/9/2025.Additional information was requested, and the following was obtained: what was the grade of hemorrhoids? 2-3 were the hemorrhoids circumferential or in specific quadrants? please describe.Circumferential what is the surgeon¿s experience with the pph procedure? 1000+ procedures over the last two years.What is the surgeon¿s experience with the pph device? very experienced where in the green gap setting scale was the indicator located prior to firing (low-b, middle-b, or high-b)? did the surgeon wait 30 seconds after closing the device and then 20 seconds after firing? yes was a complete washer transection confirmed? yes were there any issues noted with staple formation at any point throughout the care of the patient? if so, please describe the shape and location.No what is the current patient status?.
|
| |
|
Search Alerts/Recalls
|