| Model Number |
SM822 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Unraveled Material (1664)
|
| Health Effect - Clinical Codes |
Wound Dehiscence (1154); Inflammation (1932); Post Operative Wound Infection (2446)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant medical product: sm-691, sm-691 biosyn* 4-0 und 75cm c13 x36 (lot#d4f2915fy) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that following a breast reduction plasty procedure, issues arose with the sutures 3.0 and 4.0, specifically the sm-822 and sm-691 models.Approximately six weeks post-operation, pus pimples developed at the locations of subcutaneous suture knots.Additionally, stitches emerged from the scars resembling worms, occurring several weeks after the operation.Thread fragments from the subcutaneous and cutaneous skin layer emerge from the skin, leading to inflammation and in some cases also to suppuration.It was also identified that the suture line did not hold.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted that the returned photo noted that a suture fragment was observed in the provided image.It was reported that the suture line did not hold.Thread fragments from the subcutaneous and cutaneous skin layer emerge from the skin.The reported issue were confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that following a breast reduction plasty procedure, issues arose with the biosyn sutures 3.0 and 4.0, specifically the sm-822 and sm-691 models.Approximately 6 weeks post-operation, pus pimples developed at the locations of subcutaneous suture knots.Additionally, stitches emerged from the scars resembling worms, occurring several weeks after the operation.Thread fragments from the subcutaneous and cutaneous skin layer emerge from the skin, leading to inflammation and in some cases also to suppuration.It was also identified that the suture line did not hold.
|
| |
|
Search Alerts/Recalls
|
|
|