• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® PPT¿ PLASMA PREPARATION TUBES K2E (EDTA) 9.0 MG; BLOOD/PLASMA COLLECTION DEVICE FOR DNA TESTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® PPT¿ PLASMA PREPARATION TUBES K2E (EDTA) 9.0 MG; BLOOD/PLASMA COLLECTION DEVICE FOR DNA TESTING Back to Search Results
Catalog Number 362791
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/13/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported during use of bd vacutainer® ppt¿ plasma preparation tubes k2e (edta) 9.0 mg, 5 tubes exhibited poor plasma (floaties in the plasma).There was no health impact or consequences reported.
 
Additional Manufacturer Narrative
Investigation summary: bd had not received any samples for investigation.100 retained samples were visually inspected with no issues observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.Complaints for sample quality were not under statistical control for the month of january 2025; additional testing may be required: further clinical testing could not be conducted as the tubes were expired at the time of review.Clinical evaluation cannot be conducted on expired tubes.This complaint has not been confirmed for the indicated failure mode: poor plasma.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.Bd quality will continue to monitor sample quality complaints.
 
Event or Problem Description
It was reported during use of bd vacutainer® ppt¿ plasma preparation tubes k2e (edta) 9.0 mg, 5 tubes exhibited poor plasma (floaties in the plasma).There was no health impact or consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® PPT¿ PLASMA PREPARATION TUBES K2E (EDTA) 9.0 MG
Common Device Name
BLOOD/PLASMA COLLECTION DEVICE FOR DNA TESTING
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
matt j. connor
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key21376409
Report Number1917413-2025-00112
Device Sequence Number20988571
Product Code PJE
UDI-Device Identifier50382903627917
UDI-Public(01)50382903627917(11)240111(17)250131(10)4011394
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K972075
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date01/31/2025
Device Catalogue Number362791
Device Lot Number4011394
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/14/2025
Supplement Date Received by Manufacturer05/21/2025
Initial Report FDA Received Date02/13/2025
Supplement Report FDA Received Date06/18/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-