MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL CEREPAK FREEFORM; CEREPAK FREEFORM XTRASOFT DETACHABLE COIL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL CEREPAK FREEFORM; CEREPAK FREEFORM XTRASOFT DETACHABLE COIL SYSTEM Back to Search Results
Catalog Number XCR120410
Medical Device Problem Codes Separation Failure (2547); Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/03/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, gender, race, and ethnicity were not provided.Section d.2b: procode is krd/hcg.Based on complaint information, the device was not available to be returned for analysis.A review of manufacturing documentation associated with this lot: (31218776) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no internal actions related to device manufacture or inspection.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.With the limited information available and without the product available for analysis, the reported issue documented in the complaint could not be confirmed.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.The exact cause of the event could not be conclusively determined; however, it is possible that circumstances of the procedure and / or device manipulation / interaction may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered later.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by johnson & johnson medtech, or its employees that the report constitutes an admission that the product, johnson & johnson medtech, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Event or Problem Description
The healthcare professional reported that during an endovascular embolization procedure, two (2) 4mm x 10cm cerepak freeform xsft (xcr120410) from the same lot number: (31218776) did not detach with the detachment handle (mdh1 / 102109430) or via manual break.There was no report of any negative patient impact.On 05-feb-2025, additional information was received.Per the information the procedure was a vessel sacrifice targeting the left middle meningeal artery.The vessel measured 3.5mm.Both coils were confirmed to be from the same lot: (31218776).The lot number of the detachment handle is 102109430.The same handle was used with both coils, and it is not available for return.When the coils were removed, they were not stretched; both coils were detached, one had the detachment wire still on it.The concomitant microcatheter was an excelsior® sl-10® microcatheter (stryker).The procedure was completed using penumbra smart coil.There was no negative patient impact and no delay in the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CEREPAK FREEFORM
Common Device Name
CEREPAK FREEFORM XTRASOFT DETACHABLE COIL SYSTEM
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CERENOVUS, INC.
calle circuito int norte 1820
parque industrial salvacar
juarez 32574
MX   32574
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
303552-6892
MDR Report Key21376556
Report Number3008114965-2025-00112
Device Sequence Number12917738
Product Code KRD
UDI-Device Identifier10886704083689
UDI-Public(01)10886704083689(11)240104(17)261231(10)31218776
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K220040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberXCR120410
Device Lot Number31218776
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/03/2025
Initial Report FDA Received Date02/13/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/04/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EXCELSIOR® SL-10® MICROCATHETER (STRYKER).; MECHANICAL DETACHMENT HANDLE.
Patient Age66 YR
Patient SexMale
Patient Weight102 KG
-
-