Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, gender, race, and ethnicity were not provided.Section d.2b: procode is krd/hcg.Based on complaint information, the device was not available to be returned for analysis.A review of manufacturing documentation associated with this lot: (31218776) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no internal actions related to device manufacture or inspection.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.With the limited information available and without the product available for analysis, the reported issue documented in the complaint could not be confirmed.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.The exact cause of the event could not be conclusively determined; however, it is possible that circumstances of the procedure and / or device manipulation / interaction may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered later.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by johnson & johnson medtech, or its employees that the report constitutes an admission that the product, johnson & johnson medtech, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
|
The healthcare professional reported that during an endovascular embolization procedure, two (2) 4mm x 10cm cerepak freeform xsft (xcr120410) from the same lot number: (31218776) did not detach with the detachment handle (mdh1 / 102109430) or via manual break.There was no report of any negative patient impact.On 05-feb-2025, additional information was received.Per the information the procedure was a vessel sacrifice targeting the left middle meningeal artery.The vessel measured 3.5mm.Both coils were confirmed to be from the same lot: (31218776).The lot number of the detachment handle is 102109430.The same handle was used with both coils, and it is not available for return.When the coils were removed, they were not stretched; both coils were detached, one had the detachment wire still on it.The concomitant microcatheter was an excelsior® sl-10® microcatheter (stryker).The procedure was completed using penumbra smart coil.There was no negative patient impact and no delay in the procedure.
|