|
MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3507225BC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
10/24/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: capsular contracture and rupture.D4: udi: as the serial number for the device involved in the event was not provided, the full udi is currently not available.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a 51-year-old female patient underwent a primary breast augmentation with a 225cc mentor memorygel breast implant and experienced a rupture and capsular contracture (baker grade unknown) on the right side post-operatively, which was diagnosed with an office visit.As a result, the patient underwent explantation on (b)(6) 2025 and replaced with a 225cc mentor memorygel breast implant (catalog number 3502251bc) with serial number (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
On february 19, 2025, the evaluation for the device was completed.Device evaluation summary: visual analysis of the returned device revealed that the gel mod-rnd 225cc breast implant was ruptured and received in two parts.In addition, shell abrasion was observed at the edges of the rupture.These findings are consistent with normal wear of the device.Shell abrasion suggests in-vivo folding or creasing of the device, which may be caused by continuous and sustained stresses to the device, such as an excessively small breast pocket or the folding or wrinkling of the shell within the breast pocket.In some cases, breast implants may also experience normal wear over time.Regarding the reported capsular contracture condition, there are not enough details to determine the factors that may have caused this condition.Capsular contracture in the patient¿s breast might be the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Further, capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet the american society of plastic surgeons recommends and encourages member surgeons to always submit breast implants, capsule, and effusion to pathology for examination.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria, and that normal wear may have caused the observed defect.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|