• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH STARDRIVE SCREWDRIVER SHAFT T25/QC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH STARDRIVE SCREWDRIVER SHAFT T25/QC Back to Search Results
Catalog Number 314.119
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
 
Event or Problem Description
It was reported that on an unknown date, post-op inspection revealed that several screwdrivers were stripped and needed replacement.There was no patient consequences.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history review (dhr): the device lot number is unknown; therefore, a device history review could not be performed.If more information become available, the record will be re-assessed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STARDRIVE SCREWDRIVER SHAFT T25/QC
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key21378746
Report Number8030965-2025-01588
Device Sequence Number17106007
Product Code HXX
UDI-Device Identifier10886982188564
UDI-Public(01)10886982188564
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number314.119
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/02/2025
Supplement Date Received by Manufacturer03/11/2025
Initial Report FDA Received Date02/13/2025
Supplement Report FDA Received Date03/12/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
3.5MM LCP PLATE 7 HOLES 98MM
Patient SexUnknown
-
-