| Model Number |
S204 |
| Medical Device Problem Code |
Failure to Capture (1081)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/31/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this pacemaker system exhibited intermittent loss of capture (loc) on the right ventricular (rv) lead.The involved lead is a non-boston scientific and has been implanted since 1990.Technical services (ts) recommended further evaluation of this patient and lead.Reprogramming was performed and thresholds were increased to 3 v.No adverse patient effects were reported.This device system remains in service.
|
| |
|