| Model Number |
LS1020 |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/16/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, prior to use, both of the handles' tips did not open and also not closing.The devices were cleaned after each use.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: open sealer lf4418 curved large jaw - lf4418, lot# unknown.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection found the device showed evidence of excessive wear.Functional testing noted that the device's jaw opening and closing mechanism was functioning properly.The weld integrity of the device was inspected and was found to be within specification.The knife cut of the device was tested on a silicone test strip with acceptable results.The device's knife blade retracted and advanced properly.The heel gap of the device was measured and it was found to be within specification.The device intermittently alarmed when activation attempts were made.The device was disassembled and there was evidence of fluid ingress in the device and on the pcb board.Corrosion was noted on the connection point of the blue wire.It was reported that jaws were difficult to open and close.The reported issues were not confirmed.The most likely cause could not be established from the information available.The evaluation detected an unreported condition: fluid ingress.The product analysis noted evidence that the device was not used as intended.The manufacturing re cords for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: keep the instrument jaws clean.Build up of eschar may reduce sealing or cutting effectiveness.Wipe jaw surfaces and edges with a wet gauze pad as needed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|