ETHICON INC. PROLENE MESH 3INX6IN; MESH, SURGICAL, POLYMERIC Back to Search Results
Catalog Number PMII
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/04/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.D4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.H6 component code: g07002.
 
Event or Problem Description
It was reported that a patient underwent an unspecified procedure on (b)(6) 2024 and mesh was used.It was very thin with the rolled edges and it was very unsatisfactory and poor-quality mesh.There were no adverse patient consequences reported.No further information was provided.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3 investigation summary- the product was returned to ethicon for evaluation.One open pouch and one folder containing a piece of mesh were returned for analysis.Product code: pmii.During the visual inspection of the mesh, it was noted that the mesh was cut and showed signs of body fluids.In addition, the other section of the mesh was not returned for evaluation.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.Per the conditions of the returned sample, no conclusion could be reached as to what caused the reported complaint.As part of ethicon quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.A manufacturing record evaluation was performed for the finished device lot number, and no related non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROLENE MESH 3INX6IN
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km 8.3
san lorenzo PR 00754
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key21381225
Report Number2210968-2025-01664
Device Sequence Number15522319
Product Code FTL
UDI-Device Identifier10705031049024
UDI-Public(01)10705031049024(17)280331(10)V3001
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K180829
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberPMII
Device Lot NumberV3001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/20/2025
Supplement Date Received by Manufacturer02/13/2025
Initial Report FDA Received Date02/13/2025
Supplement Report FDA Received Date03/13/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-