| Catalog Number |
3910-500-471 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/14/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the eyelet broke during the procedure.All pieces were retrieved from the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Three attempts have been made to retrieve the device for evaluation, and it is yet to be received in-house.No further attempts to retrieve the device are required.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: eyelet broken.Probable root cause: application: excessive force or leveraging of the anchor against the bone.User fully seats eyelet anchor onto inserter which eliminates interference fit.Inadequate assessment of bone quality, pilot hole not prepared.The reported failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that the eyelet broke during the procedure.All pieces were retrieved from the patient.
|
| |
|