MAUDE Adverse Event Report: ETHICON INC. SFXSPI MCL+ UNI UD 12IN 3-0 SA PS-1 PMP; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. SFXSPI MCL+ UNI UD 12IN 3-0 SA PS-1 PMP; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID Back to Search Results
Catalog Number SXMP1B101
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/15/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3 analysis summary: the product was returned to ethicon for evaluation.One needle-suture piece outside its packaging was received for analysis.Product code sxmp1b101.Additionally, a photo was provided, and with the same condition as the received sample.Upon visual inspection of the returned sample, the section of the loop was not received for evaluation.The extreme was noted to be cut probably caused by a surgical instrument.Also, body fluids were observed on the strand as well.The instructions for use do contain the following caution: as with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of surgical instruments, such as needle holders and forceps.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.A manufacturing record evaluation was performed for the finished device lot number, and no related non-conformances were identified.
 
Event or Problem Description
It was reported that a patient underwent a c-section on (b)(6) 2025 and barbed suture was used.The loop failed to function.It was reported that the loop was loose when it was used in surgery.Changed to another one to complete the surgery.There were no adverse patient consequences reported.No additional information could be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SFXSPI MCL+ UNI UD 12IN 3-0 SA PS-1 PMP
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km 8.3
san lorenzo PR 00754
*   00754
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key21381286
Report Number2210968-2025-01665
Device Sequence Number8273668
Product Code GAM
UDI-Device Identifier10705031236776
UDI-Public(01)10705031236776(17)260331(10)100MB7
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
K151200
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2026
Device Catalogue NumberSXMP1B101
Device Lot Number100MB7
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/16/2025
Initial Report FDA Received Date02/13/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/23/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-