• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE RF 2 PROBES 50-S; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE RF 2 PROBES 50-S; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number 0279351250
Medical Device Problem Code Failure to Deliver Energy (1211)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/16/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device had insufficient / no energy delivered.
 
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Additional Manufacturer Narrative
According to what was reported, the device was discarded therefore, no further attempts to retrieve the device are required.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: device not working.Probable root cause: probe short (possibly due to fluid ingress), open circuit (possibly due to loose electrical connection), power output variation, wrong default setting, console error crossflow temperature mitigation actions use error.User attempts to sterilize, disinfect, clean or reuse probe.Incorrect power level set.Nonbendable probe bent with probe bender, or a bendable probe bent with anything other than the stryker rf probe bender.Attempting to use probe without completely immersing tip in a conductive irrigant.Probe handle is submerged in liquid or suction tube is cut.The reported failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported that the device had insufficient / no energy delivered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RF 2 PROBES 50-S
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
lucas wolski
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key21381347
Report Number0002936485-2025-00117
Device Sequence Number18258258
Product Code GEI
UDI-Device Identifier07613327056983
UDI-Public07613327056983
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K160050
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0279351250
Device Lot Number23349AE2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/16/2025
Supplement Date Received by Manufacturer04/01/2025
Initial Report FDA Received Date02/13/2025
Supplement Report FDA Received Date04/28/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/15/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-