Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/22/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported accessed and leakage reported, removed and split/hole observed.No other information was provided.It was reported this occurred with five devices.This report addresses all five devices.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device has not been received by the manufacturer for evaluation.Each event reported to bd is evaluated and investigated in accordance with our complaint investigation procedures.The investigation process includes, but is not limited to, evaluation of the event details provided by the complaint facility, a review of complaint history, manufacturing records and risk document where applicable, and an evaluation of the subject device when available to identify potential contributing factors.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a leak was confirmed.The products returned for evaluation were four 4 fr single lumen powerpicc solo.The investigation findings are consistent with damage accumulated through flexural fatigue.Flexural fatigue may occur due to cyclic kinking of the catheter tube in which physiological, placement, usage, and mechanical factors may gradually form a crack(s) in the catheter.The returned product sample was evaluated, and a split was observed between the 3 cm and 5 cm depth markers on three of the four samples.The three identified catheter splits contained physical features associated with material fatigue, and the characteristics observed which supported this type of failure included: damage which was circumferentially aligned.Fracture edges which were rounded and polished due to repeated material wear.'c' shaped impressions leading into the fracture site which are consistent with material buckling due to movement which caused the fracture edges to be pressed together.Overall elliptical shape to the fracture cross-section (a result of repeated kinking of the tubing).Repetitive kinking of the indwelling catheter appears to have contributed to the observed leak; however, the specific circumstances surrounding how kinking developed into a fracture within the catheter tubing were ultimately unknown.An examination of the catheter structure revealed no potential damage/defect related to manufacture of the product.This complaint will be recorded for future trending and monitoring purposes.
 
Event or Problem Description
It was reported accessed and leakage reported, removed and split/hole observed.No other information was provided.It was reported this occurred with four devices.This report addresses all four devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC SOLO2, TLS
Common Device Name
PERIPHERALLY INSERTED CENTRAL CATHETER (PICC)
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key21381575
Report Number3006260740-2025-00647
Device Sequence Number20950220
Product Code LJS
UDI-Device Identifier00801741028663
UDI-Public(01)00801741028663
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K072230
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number9295118
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/22/2025
Supplement Date Received by Manufacturer06/05/2025
Initial Report FDA Received Date02/13/2025
Supplement Report FDA Received Date06/27/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-