The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and will be evaluated.A supplemental will be submitted with evaluation results.
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a hole in the catheter was confirmed.The product returned for evaluation was one 4 fr sl powerpicc solo catheter.The investigation findings are consistent with damage accumulated through flexural fatigue.Flexural fatigue occurs due to cyclic kinking of the catheter tube in which physiological, placement, usage, and mechanical factors may gradually form a crack(s) in the catheter.The returned product sample was evaluated and a split was observed in the catheter tubing at the 4 cm depth marker.The catheter split contained physical features associated with material fatigue, and the characteristics observed which supported this type of failure included: ¿ damage which was circumferentially aligned ¿ fracture edges which were rounded and polished due to repeated material wear ¿ 'c' shaped impressions leading into the fracture site which are consistent with material buckling due to movement which caused the fracture edges to be pressed together ¿ overall elliptical shape to the fracture cross-section (a result of repeated kinking of the tubing) repetitive kinking of the indwelling catheter appears to have contributed to the observed split; however, the specific circumstances surrounding how kinking developed into a fracture within the catheter tubing were ultimately unknown.A measurement of the catheter outer diameter, inner diameter and wall thickness were taken.The catheter was found to be within specification.This complaint will be recorded for future trending and monitoring purposes.
|