• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO MANUFACTURING NETHERLANDS TECNIS ODYSSEY TORIC II IOL; LENS, MULTIFOCAL INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO MANUFACTURING NETHERLANDS TECNIS ODYSSEY TORIC II IOL; LENS, MULTIFOCAL INTRAOCULAR Back to Search Results
Model Number DRT150
Medical Device Problem Codes Device Difficult to Setup or Prepare (1487); Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/23/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information: section a-3b patient gender: male.Section a-4 patient weight: unknown as information was not provided.Section a-5 patient ethnicity: unknown as information was not provided.Section a-6 patient race: unknown as information was not provided.Section d-6b date explanted: not applicable as the iol remains implanted; therefore, not explanted.Section h-3 device evaluated by manufacturer: the complaint device was not returned for evaluation as it remains implanted.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch report will be filed.Attempts were made to contact the customer account requesting additional information regarding complaint however, to date no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
The pre-loaded drt150 model intraocular lens (iol) was injected into the patient's ocular sinister (left eye) and the lens was gorged on the outer side of the lens.The damage was clarified as a deep scratch on the peripheral edge.Due to the peripheral location of the scratch, the surgeon decided to leave the iol implanted.The doctor stated it was a little tight when he pushed the lens in.It is unknown if the incision was enlarged however, there was no medical attention, no medication was prescribed, no procedural delays, no sutures, and no vitrectomy during the procedure.Viscoelastic was used and the surgical technician is experienced with pre-loaded lenses.The lens are kept in a separate room within temperature outside the operating room (or).Patient's daily activities significantly were not affected.Patient prognosis post-procedure was reported as fully recovered.The suspect iol is not available for return as it remains implanted.No further information is available.
 
Additional Manufacturer Narrative
Corrected data: upon further review it was noted that section g4: pma/510(k) number in the initial mdr was inadvertently submitted missing the ¿p¿ before the numerical value.The complete number is p980040.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS ODYSSEY TORIC II IOL
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key21381932
Report Number3012236936-2025-000043
Device Sequence Number8595436
Product Code MFK
UDI-Device Identifier05050474811263
UDI-Public(01)05050474811263(17)271007
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDRT150
Device Catalogue NumberDRT1500155
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/23/2025
Supplement Date Received by Manufacturer12/15/2025
Initial Report FDA Received Date02/13/2025
Supplement Report FDA Received Date01/09/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/07/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-