| Model Number |
MI1250 SYNCHRONY 2 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Skin Tears (2516)
|
| Date of Event |
01/23/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user has been explanted due to an extrusion of the electrode in the external auditory canal.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: device investigation did not reveal any device defect or damage, which has been present whilst implanted.Mechanical damages found during investigation are attributable to the removal surgery.According to the information received from the field the device was explanted due to an extrusion of the electrode into the external ear canal.Further a complete insertion was not achieved at implantation surgery with six channels remaining extra-cochlear.This is a final report.
|
| |
|
Event or Problem Description
|
|
The user has been explanted due to an extrusion of the electrode in the external auditory canal.
|
| |
|