| Catalog Number |
520007P |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the temporary pacing electrode catheter balloon would not inflate before case started.No impact to patient and no medical intervention was required.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was unconfirmed.Visual evaluation of the returned sample noted one opened (with original packaging), used temporary pacing electrode catheter.Visual inspection of the sample noted using (in-house) syringe inflated balloon with 1.5cc air without difficulty.Allowed to rest with no leaks evident.The reported event was not confirmed.No root cause could be found because the reported event was unconfirmed.Dhr is not required since the reported failure was unconfirmed.The reported event is unconfirmed a labeling review is not required.Correction: d, e, h udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the temporary pacing electrode catheter balloon would not inflate before case started.No impact to patient and no medical intervention was required.
|
| |
|