• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 BARD® LUBRICATH® FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 BARD® LUBRICATH® FOLEY CATHETER Back to Search Results
Catalog Number 365716
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/31/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the balloon from the foley catheter popped inside their patient.This happened twice.There was no patient harm, and the customer had the defective product to return.
 
Event or Problem Description
It was reported that the balloon from the foley catheter popped inside their patient.This happened twice.There was no patient harm, and the customer had the defective product to return.
 
Additional Manufacturer Narrative
The reported event was unconfirmed because the devices meet the specification.The product had not caused the reported failure.A potential root cause for this failure mode could be low latex viscosity, short latex dwell time, latex dip speed out too slow causing thin sac.A review of the device history record did not show any problems or condition that would have contributed to the reported issue.The instructions for use were found adequate and state the following: insert foley catheters only for appropriate indications and leave in place only as long as needed.Proper techniques for urinary catheter insertion: perform hand hygiene immediately before and after insertion.Insert urinary catheters using aseptic technique and sterile equipment.Use the smallest foley catheter possible, consistent with good drainage.Document the indications for catheter insertion, date and time of catheter insertion, individual who inserted catheter, and date and time of catheter removal in the patient record.Proper techniques for urinary catheter maintenance: secure the foley catheter.Use the statlock foley stabilization device if provided.Maintain a closed drainage system by utilizing pre-connected, sealed catheter-tubing junctions.Maintain unobstructed urine flow and keep the catheter and collection tube free from kinking.Keep the collection bag below the level of the bladder or hips at all times.Empty the collection bag regularly using a separate, clean collection container for each patient.Generally, drainage is accomplished by inserting the catheter through the urethra and into the bladder.However, drainage is sometimes accomplished by suprapubic or other placement of the catheter, such as nephrostomy tract.Correction: d, e, h.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® LUBRICATH® FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key21383900
Report Number1018233-2025-00897
Device Sequence Number8387872
Product Code EZC
UDI-Device Identifier00801741077746
UDI-Public(01)00801741077746(11)230606(17)280128(10)MYHP4288
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K910846
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number365716
Device Lot NumberMYHP4288
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2025
Supplement Date Received by Manufacturer05/09/2025
Initial Report FDA Received Date02/14/2025
Supplement Report FDA Received Date05/13/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/06/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-