| Catalog Number |
8065000334 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Intraocular Pressure Decreased (4468)
|
| Date of Event |
01/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Event or Problem Description
|
|
A non healthcare professional reported that the fluidics module displayed an error code, aspiration could not function, valved cannulas were leaking and eye was too soft which was stabilized by perfluropropane (c3f8) and a syringe with thirty gauge needle during surgery.The procedure type was unknown.The procedure was completed.It was unknown about how the procedure was completed.There was no patient impact.Additional information has been requested.None received till date.This report pertains to entry system involved in reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in h.6.And h.11.A sample was not received at the manufacturing site for evaluation for the report.Therefore, the condition of the product could not be verified.A review of the device history records traceable to the reported lot number indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.A sample was not received at the manufacturing site and the device history record review of the lot number provided indicated the product was processed and released according to the product¿s acceptable criteria, therefore; the root cause for the customer complaint issue cannot be determined.The exact root cause for this complaint is unknown; therefore, specific action with regards to this complaint cannot be taken.All trocars are 100% inspected.This inspection includes evaluation for damaged valves.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Search Alerts/Recalls
|