• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL SA X-SERIES RECIPROCATING SAW ATTACH; SAWING POWER TOOL ATTACHMENT, RECIPROCATING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL SA X-SERIES RECIPROCATING SAW ATTACH; SAWING POWER TOOL ATTACHMENT, RECIPROCATING Back to Search Results
Catalog Number 89-8522-450-20
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/28/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).G2 ¿ foreign ¿ romania.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported that the motor stopped working during surgery with the reciprocating saw attachment on.Attempts have been made and no further information has been provided.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: b4, b5, d2, d4, d9, g1, g3, g6, h1, h2, h3, h6, h11.Review of the most recent repair record determined the tip was damaged/loose and the device did not rotate.Unknown what internal component was damaged/broken.The device was scrapped.Review of the device history record identified no deviations or anomalies during manufacturing.Review of complaint history identified additional similar complaints for the reported item and no additional similar complaints for the reported part and lot combination.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-SERIES RECIPROCATING SAW ATTACH
Common Device Name
SAWING POWER TOOL ATTACHMENT, RECIPROCATING
Manufacturer (Section D)
ZIMMER SURGICAL SA
3, ch. du pre fleuri
1228 plan-les-ouates
geneva CH-12 28
SZ  CH-1228
Manufacturer (Section G)
ZIMMER SURGICAL SA
3, ch. du pre fleuri
1228 plan-les-ouates
geneva CH-12 28
SZ   CH-1228
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key21392920
Report Number0008031000-2025-00033
Device Sequence Number20929801
Product Code GEY
UDI-Device Identifier00889024521315
UDI-Public(01)00889024521315(11)230308(10)5018354
Combination Product (Y/N)N
Initial Reporter CountryRO
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number89-8522-450-20
Device Lot Number5018354
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2025
Supplement Date Received by Manufacturer07/01/2025
Initial Report FDA Received Date02/18/2025
Supplement Report FDA Received Date07/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/08/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-