|
Additional information: d9, g3, h3, h6.The complaint allegation was confirmed.One unpackaged ar-3350-4031 4k arthroscope, 30°, 4 mm x 152.5 mm, serial/batch number (b)(6), was received for evaluation.Visual inspection revealed nicks and damage on the tip of the scope, indicating possible interaction with another instrument during use or mishandling.Functional testing was conducted due to the damage.The most likely cause of the reported failure is user mishandling.The instrument was manufactured on october 9, 2023.
|