|
Additional information: d9, g3, h3, h6.Complaint allegation is confirmed.One unpackaged radiolucent targeting arm, 1268-100, batch/serial (b)(6), was received for investigation.Upon visual evaluation, it was noted that the threads were damaged.Additionally, the device had surface damage along the body: fading, chipping and nicks.Functional was not performed due to the damage to the devices.The most likely reason for the reported failure is the wear and tear damage incurred over repeated usage.The manufacturing date is 2023.
|