• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED CELLEX PHOTOPHERESIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED CELLEX PHOTOPHERESIS SYSTEM Back to Search Results
Catalog Number CLXSP-I
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/31/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction pressure dome leak.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot n125 was conducted.There were no non-conformances associated with this lot.This lot met all release requirements.A review of kit lot n125 shows no trends.Trends were reviewed for complaint category, pressure dome leak.No trends were detected for this complaint category.At the time of this report, the analysis of the returned photographs is still in process.A final report will be filed when the analysis is complete.(b)(4).
 
Event or Problem Description
The customer contacted mallinckrodt to report they experienced a pressure dome leak with their cellex photopheresis kit ("kit") during an extracorporeal photopheresis (ecp) treatment.The customer reported that the return pressure dome leaked during treatment after 100ml of whole blood had been processed.The ecp treatment was aborted and no residual blood within the kit was returned to the patient.The customer reported the patient was in stable condition.The customer returned photographs for evaluation.
 
Additional Manufacturer Narrative
Photographs were returned for evaluation.The complaint kit and smart card were not returned for evaluation.Examination of the customer provided photograph confirmed the leak from the pressure dome.A material trace of the pressure dome housing assembly and its components used to build lot n125 did not find any non-conformances.The device history record (dhr) review did not result in any related non-conformances.This lot passed all lot release testing.The reported pressure dome leak is verified based on the photographs provided; however, the root cause for the pressure dome leak could not be determined based on the available information.No further action is required at this time.This investigation is now complete.(b)(4) 13 mar 2025.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Common Device Name
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
blanchardstown, dublin D15 T X2V
EI  D15 TX2V
Manufacturer (Section G)
HARMAC MEDICAL PRODUCTS
i.d.a. business park
castlerea roscommon,
EI  
Manufacturer Contact
mark wendelken
440 route 22 east
suite 302
bridgewater, NJ 08807
MDR Report Key21395858
Report Number3013428851-2025-00017
Device Sequence Number15752866
Product Code LNR
UDI-Device Identifier00705030000005
UDI-Public(01)00705030000005(10)N125(17)260201
Combination Product (Y/N)Y
Initial Reporter CountryUK
PMA/510(K) Number
P860003
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2026
Device Catalogue NumberCLXSP-I
Device Lot NumberN125
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/31/2025
Supplement Date Received by Manufacturer02/13/2025
Initial Report FDA Received Date02/18/2025
Supplement Report FDA Received Date03/13/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/27/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-