Visual inspection: during the investigation, scratches on the outer housing of the progav 2.0 were determined.Permeability test: a permeability test has shown that all components are permeable.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that both valves operate within the permissible tolerance in their respective relevant positions.Adjustment test: the progav 2.0 was tested and is not adjustable.Braking force and brake function test: the brake functionality test has shown that the brake function operates as expected; however, the breaking force cannot be measured due to the non-adjustability of the valve.Internal inspection: after dismantling of the valves, deposits were found in progav 2.0.Results: based on our investigation results, we can determine a non-adjustability at the progav 2.0.The visible deposits in the valve have led to the functional impairment.Deposits of natural substances found in the body, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of proteins can affect the integrity of the valve.Furthermore, we can determine no functional deviations or other abnormalities in the shuntassistant 2.0.It met its specification.The examination of the return shipment is completed with the creation of this report.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|