• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG M.BLUE 10 VALVE; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG M.BLUE 10 VALVE; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX802T
Medical Device Problem Codes Infusion or Flow Problem (2964); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 01/14/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Visual inspection during the investigation, scratches on the outer housing of the valves , but no significant deformations or damage was determined.Permeability test a permeability test has shown that all components are permeable.Computer controlled test to investigate the claim of over- drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in vertical/ or both the horizontal as well as the vertical positions.The results show that the m.Blue is operating within the accepted tolerance in the horizontal position, but not within the specified tolerances in the vertical position.An slower outflow could be determined.Adjustment test the m.Blue valve was tested and is not adjustable throughout the normal range.Braking force and brake function test the brake functionality test has shown that the brake function is operational, however the braking force cannot be measured due to the non-adjustability of the valve.Internal inspection after dismantling of the valve, deposits were found.Results based on our investigation results, we can determine a slower outflow and adjustment difficulties.The determined deposits can be named as the cause for the functional deviation.Proteins in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.Based on our investigation results, we can determine deposits.The deposits had no affect to the technical properties at the time of the investigation.The cause of the aforementioned functional impairment is not known to us at the time of the examination.Proteins in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Event or Problem Description
It was reported that an m.Blue (#fx802t) was implanted during a procedure performed on (b)(6) 2024.According to the complainant, the shunt system caused an overdrainage and had adjustment difficulties.The patient underwent a revision procedure on (b)(6) 2025.No patient complications were reported as a result of the revision procedure.The complained device was returned to the manufacturer for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M.BLUE 10 VALVE
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key21396732
Report Number3004721439-2025-00031
Device Sequence Number9327914
Product Code JXG
UDI-Device Identifier04041906504410
UDI-Public4041906504410
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K192266
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFX802T
Device Catalogue NumberFX802T
Device Lot Number20071280
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/21/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/21/2025
Initial Report FDA Received Date02/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/07/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age8 YR
Patient SexFemale
Patient Weight22 KG
-
-