|
Updated field: b4, g3, g6, h2, h3, h6 (component code, investigation findings, investigation conclusions, type of investigation), h11.A getinge field service engineer (fse) evaluated the device and several faults were detected.Fse replaced pressure sensors x1 and x2 faulty (0682-00-0076-01), drive assembly (0104-00-0018), fill manifold (0104-00-0023).The following was also missing: support (iv pole) (0436-00-0199), power cable (0012-00-0886-02), bag with doppler (0154-01-0001).The material to replace and replenish was ordered.On 5-02-2025 replacement of identified components and calibration of all equipment.Verification, preparation of service protocol and functional tests.Operational equipment.Unit returned for clinical use.
|