| Model Number |
33410ON |
| Medical Device Problem Codes |
Break (1069); Insufficient Device Problem Information (3190); Appropriate Device Problem Term/Code Not Available (3191); Sharp Edges (4013)
|
| Health Effect - Clinical Codes |
Perforation (2001); Insufficient Information (4580)
|
| Date of Event |
02/05/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The affected device has been requested for investigation by the manufacturer.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted unsolicited.Cross reference complaint id: (b)(4).This event is filed under internal complaint id (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the atraumatic grasper created a tear/hole on the stomach, and it was not being pulled on.The surgeon said that she would have needed to suture the holes, had that part of the stomach remained inside the patient.Fortunately, it was the part of the stomach that was removed from the body.The surgeon mentioned that it happened in another case a week or two ago.Cross reference mdr: 9610617-2025-00241.
|
| |
|
Additional Manufacturer Narrative
|
|
Per investigation findings by the manufacturing site:
during the analysis of the similar case it was found that the pull rod was significantly bent in the upper and lower areas.As the pull rod must be straight due to its design, this deformation is considered atypical.Due to the material properties and mechanical stability of the component, it can be assumed that the bending was caused by excessive force during use.Such deformation is not to be expected under normal conditions of use.Based on the results of the investigation, the documented product quality, and the previous complaint history, no product-related defect can be identified.The cause of the reported injury cannot therefore be attributed to a defect in the item.This event is filed under internal complaint id (b)(4).
|
| |
|