• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. SWIVELOCK, SP PEEK 5.5MM; NON-DEGRADABLE FIXATN FASTENER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. SWIVELOCK, SP PEEK 5.5MM; NON-DEGRADABLE FIXATN FASTENER Back to Search Results
Model Number SWIVELOCK, SP PEEK 5.5MM
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/29/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On (b)(6) 2025, it was reported by a sales representative via email that the tip eyelet on three ar-2323psp self-punching peek swivelock were not properly set on the drivers, and they broke during insertion.It seemed like there was slack in the sliding suture causing the eyelid to flip sideways when punching in.This was discovered during a procedure on (b)(6) 2025.Additional information requested on 2/7/2025.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6, h11.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause is attributed to misuse due to prying/leveraging the device during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWIVELOCK, SP PEEK 5.5MM
Common Device Name
NON-DEGRADABLE FIXATN FASTENER
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key21399121
Report Number1220246-2025-00511
Device Sequence Number13141321
Product Code MAI
UDI-Device Identifier00888867384491
UDI-Public(01)00888867384491(11)231227(17)281130(10)15186153
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K222263
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSWIVELOCK, SP PEEK 5.5MM
Device Catalogue NumberAR-2323PSP
Device Lot Number15186153
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/04/2025
Supplement Date Received by Manufacturer08/24/2025
Initial Report FDA Received Date02/18/2025
Supplement Report FDA Received Date08/25/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/27/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-