| Model Number |
DISPOSABLE KIT, KNEE FIBERTAKS |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2025, a sales representative reported via email that an ar-3710 disposable drill guide kit and (qty.2) of an ar-3770sp hybrid knee fibertak anchor failed while implanting using the drill guide kit.No further information was provided.This was discovered during a procedure, with no reported patient harm.Additional information was received on 02/11/2025: the ar-3710 disposable drill guide kit did not remove enough bone.It did not produce debris.The (qty.2) of an ar-3770sp hybrid knee fibertak anchor would not insert into the bone and would just bend when being hammered in.No instrument or implant broke during this issue.The case was completed using a 4.75 swivelock mpfl kit.There was no case delay or added anesthesia.There was no adverse effects on the patient reported.This occurred during a mpfl procedure on (b)(6) 2025.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to misaligned insertion.
|
| |
|