BECKMAN COULTER, INC. AU680 CLINICAL CHEMISTRY ANALYZER; ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE
|
Back to Search Results |
|
| Model Number |
AU681-10E CHEMISTRY ANALYZER AU680 WITH ISE |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A beckman coulter customer technical support (cts) specialist assisted customer over the phone.The cts suggested part replacements and was unable to resolve the issue over the phone.A beckman coulter field service engineer (fse) was dispatched to the customer site to evaluate the au680 clinical chemistry analyzer.The fse inspected the analyzer and found the spiral mix bar fit loose on the mixer assembly unit.The fse replaced the spiral mix bar to resolve the issue.Performed system verification successfully without issues.The system returned to normal operation.Section a2, a4, and a5: information not provided by customer.The beckman coulter identifier is (b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported obtaining two (2) false low ise (ion selective electrode) patient results which include sodium (na), chloride (cl), and potassium (k) on their au680 clinical chemistry analyzer.The erroneous results were not reported outside the laboratory.The customer did not provide patient data or demographics, and the number of affected patient samples is unknown.There was no report of change to treatment, injury, or death associated to this event.
|
| |
|