| Catalog Number |
MZ9266 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd maxzero multi-fuse extension set with needleless connector was breaking the following information was received by the initial reporter with the following verbatim it was reported by the customer that here were cracks on the luer lock portion of trifuse while in use on patient.
|
| |
|
Event or Problem Description
|
|
No additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Two samples model mz9226 under (b)(4) were returned for investigation together.The sets were examined for defects and abnormalities.Both spin collars were cracked and one of the spin collars was separated from the male luer.No other defects or abnormalities were observed.The customer complaint was verified and quality notification was sent to the supplier.The assembly process was thoroughly reviewed and found no evidence indicating this issue stems from an assembly-related factor.Review of device history records confirmed that no issues were identified during the manufacturing of the provided lots.Additionally, no non-conformance material reports were filed during the locknuts¿ manufacture period concerning this issue.An investigation into the maintenance work order history around the time of manufacturing revealed that no maintenance activities were related to or caused the reported issue.The most probable root cause of the luer cracking is due to the alcohol from the antiseptic swab.
|
| |
|
Search Alerts/Recalls
|