| Catalog Number |
8065000334 |
| Medical Device Problem Code |
Suction Failure (4039)
|
| Health Effect - Clinical Code |
Intraocular Pressure Decreased (4468)
|
| Date of Event |
01/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A non healthcare professional reported that the fluidics module displayed an error code, aspiration could not function, valved cannulas were leaking and eye was too soft which was stabilized by perfluropropane (c3f8) and a syringe with thirty gauge needle during surgery.The procedure type was unknown.The procedure was completed.It was unknown about how the procedure was completed.There was no patient impact.Additional information has been requested.None received till date.This report pertains to cassette involved in reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The sample was not received at the investigating site for this complaint report; visual inspection or functional testing could not be conducted.If a sample is returned at a later date, the investigation will be reopened and the sample evaluated.The root cause of the customer's complaint could not be established as a sample has not been received.Without a sample, it is not possible to isolate the root cause.As the root cause is unknown, the relationship, if any, of the device to the reported incident cannot be determined.After the investigation of this complaint, it has been determined that no action will be taken at this time as a sample was not returned.No adverse trends have been observed associated with the reported product and event.Quality assurance will continue to monitor customer complaints via the complaint review meetings and will take action for any future occurrences as is deemed necessary.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Search Alerts/Recalls
|