It was reported that the balloon would not inflate, and the sleeve of the device was inadvertently left in the patient and had to be surgically removed.The patient was referred for a vascular procedure post mitral valve area (mva) for vascular arterial thrombus.A flap dissection was found in the tibioperoneal (tp) trunk.During the initial procedure on (b)(6) 2024, a 3.00 x 38mm synergy xd drug-eluting stent was advanced for treatment, however the balloon would not inflate.The device was removed, and a different stent was deployed without issue.However, the patient was brought back to the operating room (or) for thrombus formation and exploratory surgery.A.Surgery revealed a foreign body of a piece of plastic cover present on all coronary stents and balloons along with the a stylet, found in the popliteal artery.It is believed that the plastic cover/sleeve was left in the body during withdrawal of the stent delivery system.No fragments were left inside the patient after the surgery, and the patient was expected to fully recover.
|