BOSTON SCIENTIFIC CORPORATION LEVEEN STANDARD; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number M001262150
Medical Device Problem Code Retraction Problem (1536)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/23/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the tines of the needle were difficult to retract.This 3.5/15/25 ph leveen standard needle electrode was selected for a radiofrequency ablation procedure to treat a possible tumor.After successfully deploying and performing ablation six times using the device, it was attempted to retrieve the needle, but this was unsuccessful.More force was applied without success; therefore, they captured the needle by pushing non-ionic water into the side of the needle to clear any obstacles.The procedure was completed, and the needle was removed.Once outside the body, it was noted that there were particles stuck in the needle, and once removed, the tines could be retracted normally.There were no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEEN STANDARD
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key21401271
Report Number2124215-2025-06448
Device Sequence Number21226566
Product Code GEI
UDI-Device Identifier08714729349594
UDI-Public08714729349594
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K140495
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM001262150
Device Catalogue NumberM001262150
Device Lot Number0035158812
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/23/2025
Initial Report FDA Received Date02/18/2025
Date Device Manufactured10/28/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age78 YR
Patient SexMale
Patient Weight60
-
-