• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. CEMENTED FINNED TIB. TRAY SZ 1DF/2T; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. CEMENTED FINNED TIB. TRAY SZ 1DF/2T; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 200-06-12
Medical Device Problem Codes Naturally Worn (2988); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Scar Tissue (2060); Synovitis (2094); Osteolysis (2377)
Date of Event 11/16/2017
Type of Reportable Event Serious Injury
Event or Problem Description
As reported via legal documentation a 59 y/o patient had a right knee replacement on (b)(6) 2011.Approximately 6 years after the initial procedure the patient had a right knee revision on (b)(6) 2017 due to severe articular surface wear.
 
Additional Manufacturer Narrative
The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: g3/g4, h6.The following sections were corrected: d1/d2a/d2b, d4, d6a, h4, h6.Mdr section codes updated/corrected: b, e.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, femoral loosening, and tibial loosening.Potential contributions of manufacturing, user, and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CEMENTED FINNED TIB. TRAY SZ 1DF/2T
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
MDR Report Key21401460
Report Number1038671-2025-01012
Device Sequence Number15507712
Product Code JWH
UDI-Device Identifier10885862039903
UDI-Public10885862039903
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K011976
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/21/2014
Device Catalogue Number200-06-12
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Distributor Facility Aware Date05/21/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/21/2023
Supplement Date Received by Manufacturer10/22/2009
Initial Report FDA Received Date02/18/2025
Supplement Report FDA Received Date03/12/2025
Date Device Manufactured10/22/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age59 YR
Patient SexFemale
-
-