• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® BORIC ACID SODIUM BORATE/FORMATE C&S URINE TUBE; CULTURE MEDIA, NON-PROPAGATING TRANSPORT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® BORIC ACID SODIUM BORATE/FORMATE C&S URINE TUBE; CULTURE MEDIA, NON-PROPAGATING TRANSPORT Back to Search Results
Catalog Number 364958
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/29/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that while using bd vacutainer® boric acid sodium borate/formate c&s urine tube, an unspecified number of tubes are not reseating on their automated instrument (copan wasp); caps are loose and causing spills and splashing laboratory worker.No adverse impact was reported regarding the patient or user.
 
Additional Manufacturer Narrative
D4.Medical device expiration date: unknown.H4.Device manufacture date: unknown.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported that while using bd vacutainer® boric acid sodium borate/formate c&s urine tube, an unspecified number of tubes are not reseating on their automated instrument (copan wasp); caps are loose and causing spills and splashing laboratory worker.No adverse impact was reported regarding the patient or user.
 
Additional Manufacturer Narrative
Investigation summary: bd had not received any samples for investigation.The device history records could not be reviewed as the lot number is unknown.This complaint has not been confirmed for the indicated failure mode: stopper pop off.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® BORIC ACID SODIUM BORATE/FORMATE C&S URINE TUBE
Common Device Name
CULTURE MEDIA, NON-PROPAGATING TRANSPORT
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
matt j. connor
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key21401838
Report Number1917413-2025-00132
Device Sequence Number9518804
Product Code JSM
UDI-Device Identifier50382903649582
UDI-Public(01)50382903649582
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K024240
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number364958
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/21/2025
Supplement Date Received by Manufacturer03/05/2025
Initial Report FDA Received Date02/18/2025
Supplement Report FDA Received Date04/04/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-