• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. J-VAC RESERVIOR BULB 100 CC; CATHETER TIP SUCTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. J-VAC RESERVIOR BULB 100 CC; CATHETER TIP SUCTION Back to Search Results
Catalog Number 2160
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Additional information: h6 component code: g07002 - device not returned.Additional information provided: patient status/ outcome / consequences: no, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: unknown, is the patient part of a clinical study : unknown.Attempts to obtain the device have been made.If further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported a patient underwent an unknown procedure on (b)(6) 2025 and a reservoir was used.Device does not perform suction.Additional information has been requested.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Additional information: d9.H3 evaluation: complaint sample review: one complaint sample of reservoir bulb was received for evaluation, product was inspected visually as well as functionally and in reference to the complaint details "device does not perform suction.", no negative observation was identified (video of functional test attached for reference).As per standard practice, 100% functional test and 100% visual inspection was carried out, before and after packing of finished goods, prior to the product release.There was no scope to miss such defect, at manufacturing / release stage.Documents review: during investigation of complaint, batch review of in-process and final packaging inspections, as well as the manufacturing process is performed, these products are reported to be manufactured, inspected, and packaged in conformance with customer's specifications and have been found to be acceptable within all parameters.No discrepancy as per the reported defect is observed.Retention sample review : no negative observation was found.Review of defective sample's image / video: not applicable, as there was no complaint sample image / video received for evaluation.The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
J-VAC RESERVIOR BULB 100 CC
Common Device Name
CATHETER TIP SUCTION
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
DEGANIA INDIA
251, sector-6, imt manesar
gurugram 12205 2
IN   122052
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key21403234
Report Number2210968-2025-01730
Device Sequence Number15584393
Product Code JOL
UDI-Device Identifier10705031003293
UDI-Public(01)10705031003293(11)240411(17)290331(10)J2400066
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
CL II EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number2160
Device Lot NumberJ2400066
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/28/2025
Supplement Date Received by Manufacturer02/10/2025
03/26/2025
Initial Report FDA Received Date02/18/2025
Supplement Report FDA Received Date03/07/2025
04/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/13/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-