• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. REMSTAR AUTO A-FLEX; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. REMSTAR AUTO A-FLEX; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number AH560HBT
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer was contacted in reference to the voluntary field safety notice/recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging an issue related to a remstar auto a-flex device's sound abatement foam.The patient has alleged visualization of particles.There was no report of serious or permanent harm or injury.No medical intervention was required by the patient.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Additional Manufacturer Narrative
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging visualization of particles.There was no report of medical intervention.No additional information can be requested at this time.The manufacturer was made aware of this complaint through a representative of the customer.The device was returned to the third-party service center for further evaluation.During the evaluation of the device at the third-party service center, the device was visually inspected and found no evidence of foam degradation.There was 0 error codes found.The secondary findings were observed.The third-party service center concludes that they couldn't confirm the customer's allegation and there was no visible foam degradation and unit got corrected.In box g: date received by mfg (rfb) has been updated.In box h: evaluation method code grid, evaluation results code grid, component code grid and conclusion code grid has been updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REMSTAR AUTO A-FLEX
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key21404970
Report Number2518422-2025-017934
Device Sequence Number8215300
Product Code BZD
UDI-Device Identifier00606959023925
UDI-Public(01)00606959023925
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 06/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberAH560HBT
Device Catalogue NumberAH560HBT
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/14/2022
Supplement Date Received by Manufacturer07/04/2024
Initial Report FDA Received Date02/18/2025
Supplement Report FDA Received Date06/09/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/09/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1974-2021
Patient Sequence Number1
Patient SexUnknown
-
-