Based on the available information, this event is deemed to be a reportable malfunction.Photograph, video and/or physical sample evaluation: there was a photograph associated with this case and in it, the reported defect can be seen.No unused return sample was expected.Batch record revision results: lot 4k03371 was manufactured on 14/oct/2024, in ats #2 line, with a total of (b)(4) market units (mkus).The complaints investigator performed a batch record review on 14/feb/2025, to verify if all the applicable procedures were followed and no issues were found; all the components for assembly were correct per bill of materials (bom) and all the tooling information documented was also correct, system application product (sap) material identification (id) 1703891 and manufacturing order (b)(4).Review of the batch record showed that all relevant tests required during the manufacturing process and final product release had been fulfilled and met the requirements.No discrepancy related to this issue were found within the documentation.The reported malfunction could be generated during the manufacturing process of the final line, therefore no further batch record review of the subassembly lots was required.Historical complaints review: on 14/feb/2025, complaints investigator ran a query in database in order to verify the complaints reported for the 3c05083 lot for the malfunction ¿skin barrier starter hole is defective (e.G misalignment or off center), leakage may occur (moldable only)¿ defect and no additional complaints were identified.Historical nonconformance review: on 14/feb/2025, complaints investigator ran a query in database looking for any in process nonconformance / corrective and preventive actions (capa) (s) associated to the malfunction ¿skin barrier starter hole is defective (e.G misalignment or off center), leakage may occur (moldable only)¿ defect for the lot number 3c05083 and as result, no nonconformance / corrective and preventive actions (capa) (s) for this malfunction were generated during the manufacturing process of the referenced lot.Defect rate analysis: there have only been 1 defective part confirmed to date from a lot size (b)(4) products.This represents a defect rate of only (b)(4), which is well within an appropriate acceptable quality level (aql) for this defect which should be 0.25% based on our standard operating procedure (sop).In addition, all of the in-process testing on this lot did not find a single defective unit, which confirms that the lot is unlikely to breach an aql of 0.25.This issue certainly appears to be an isolated incident, but more importantly to date, it is well within our accepted aql level for this type of failure mode or defect.Conclusions: as a photograph was available for this complaint issue, it was evaluated, and as a result, the reported malfunction for the observed product was confirmed, however, after performing the defect rate analysis for this issue, it was observed that the quantity of reportedly affected products is within the appropriate acceptable quality level (aql) for this defect.A batch record review was completed and showed that all relevant tests required during the manufacturing process and final product release had been fulfilled and met the requirements.No discrepancy related to this issue were found within the documentation.Additionally, no harm was reported due to this complaint issue, and the reported unit wasn¿t utilized.This issue will be monitored through the post market product monitoring review process, standard operating procedure (sop).To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092 manufacturing site: 9618003.
|