Model Number DVTX-8/110 AO NP
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 01/17/2025
Type of Reportable Event Malfunction
Event or Problem Description
(b)(6) reported on (b)(6) 2025 that the dvtx-8/110 ao np tip snapped off inside head of a tpeg.It was reported that the device was not being used under power as the surgeon was using the hndl-quick by hand, the broken tip was not able to be removed and the surgeon reported the screw was adequately locked and opted to leave implant with screwdriver tip insitu.The surgeon reported no time delay, however, believed there is potential patient impact if the most distal screw with broken screwdriver tip requires revision.A picture was provided, no x-rays were provided but the remainder of the device is available to be returned.(b)(6) stated on (b)(6) 2025 that the device is no longer available to be returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORX 8 DRIVER, 110MM NIPPLE TIP
Common Device Name
TORX 8 DRIVER, 110MM NIPPLE TIP
Manufacturer (Section D)
TRIMED, INC.
27533 avenue hopkins
santa clarita CA 91355
Manufacturer (Section G)
TRIMED, INC.
27533 avenue hopkins
santa clarita CA 91355
Manufacturer Contact
megan rissler
27533 avenue hopkins
santa clarita, CA 91355
MDR Report Key21405891
Report Number2031009-2025-00003
Device Sequence Number13090885
Product Code HXX
UDI-Device Identifier00842188126055
UDI-Public00842188126055
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDVTX-8/110 AO NP
Device Catalogue NumberDVTX-8/110 AO NP
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date02/18/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-