H6: code c19 - a review of the manufacturing records indicated the lot met all pre-release specifications.H6: code b20 -the device remains implanted and was therefore not available for engineering evaluation by gore.According to the gore® tag® conformable thoracic stent graft with active control system instructions for use (ifu), complications associated with use of the gore® tag® thoracic endoprosthesis may include but are not limited to endoleaks, aortic expansion (aneurysm) and reoperation.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
The following was reported to gore: on (b)(6) 2024, the patient underwent an unknown endovascular procedure utilizing a gore® tag® conformable thoracic stent graft with active control system.On (b)(6) 2025, the patient underwent a reintervention procedure utilizing two gore® tag® conformable thoracic stent graft with active control system (tgmr313115 & tgmr313120) and a gore® viabahn® vbx balloon expandable endoprosthesis to treat a proximal type 1 endoleak with growing aneurysm (size unknown).Reportedly, physician extended proximally with a tgmr313115 first over to the left common carotid artery.The device came back a little bit then physician added a tgmr313120 which sealed the endoleak perfectly, after which, physician had to put a gore® viabahn® vbx balloon expandable endoprosthesis into the left common carotid artery as a precaution as the device got a little too far across the ostium.There was no issues with blood flow during the procedure and blood flow remained efficient as the procedure ended successfully with patient being discharged the same day.Physician suspects cause of the endoleak and aneurysm growth is due to the patient's vehicular transaction accident.
|