• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ENCOR ENSPIRE SYSTEM; REUSABLE BIOPSY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ENCOR ENSPIRE SYSTEM; REUSABLE BIOPSY Back to Search Results
Catalog Number E4115
Medical Device Problem Code Application Program Problem (2880)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/23/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a biopsy procedure, the driver did not calibrate and returned to the start screen without any command.It was further reported that the pedal calibrated successfully but asked us to remove the driver after collecting the two to three fragments, furthermore, the driver calibrates correctly, however the system returns to the home screen without any commands.Reportedly, the communication issue has been occurred between the driver or foot pedal and the source of communication failure could not be identified.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a dhr and manufacturing review was not required as the device was not returned.Investigation summary: the physical device was not returned for evaluation.No photos or videos were provided for review.Therefore, the investigation is inconclusive for the alleged issue as no objective evidence was provided for review.A definitive root cause for the reported system returns to home screen issue could not be determined based upon the provided information.Labeling review: the labeling/packaging review was not required as the reported event is not associated with a labeling, packaging, or use related issue.H10: g3, h6 (method).H11: d4 (unique identifier (udi) #), h6 (result, conclusion).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a biopsy procedure, the driver did not calibrate and returned to the start screen without any command.It was further reported that the pedal calibrated successfully but asked us to remove the driver after collecting the two to three fragments, furthermore, the driver calibrates correctly, however the system returns to the home screen without any commands.Reportedly, the communication issue has been occurred between the driver or foot pedal and the source of communication failure could not be identified.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCOR ENSPIRE SYSTEM
Common Device Name
REUSABLE BIOPSY
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
DYMAX CORP. -2523003
110 marshall drive
warrendale PA 15086
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key21407711
Report Number2020394-2025-00193
Device Sequence Number20894165
Product Code KNW
UDI-Device Identifier00801741086113
UDI-Public(01)00801741086113(17)451130
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryBR
PMA/510(K) Number
K111100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberE4115
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/23/2025
Supplement Date Received by Manufacturer04/07/2025
Initial Report FDA Received Date02/19/2025
Supplement Report FDA Received Date04/18/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-