| Model Number |
AA07 |
| Medical Device Problem Codes |
Overheating of Device (1437); Loss of Power (1475)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/22/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Repair request initiated for device with the report of overheating and device got seized.It was reported that there was no patient involvement.Repair request escalated to a product event due to return in less than 90 days from purchase or repair.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis:evaluation could not confirm the reported malfunction of overheating the device was tested and the maximum temp erature was measured to be 85°f.It was also noted that laser markings are faded.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|