| Model Number |
MI1200 SYNCHRONY PIN |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected after a fall.The user noticed a sudden change in hearing with the device but mentioned that he did not hit his head.The user will be reimplanted, no date has been set yet for the surgery.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected after a fall.The user noticed a sudden change in hearing with the device but mentioned that he did not hit his head.The user will be reimplanted, no date has been set yet for the surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information, a damage to the active electrode due to the reported mechanical impact appears very likely.However, to determine an exact root cause, device investigation would be necessary.Re-implantation is considered but no date has been scheduled yet.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected after a fall.The user noticed a sudden change in hearing with the device but mentioned that he did not hit his head.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information, a damage to the active electrode due to the reported mechanical impact appears very likely.However, to determine an exact root cause device investigation would be necessary.Re-implantation is considered but no date is available yet.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: overload fractures in the active electrode which are consistent with an external mechanical impact were determined to be the root cause of device failure.The problems given in the recipient report appear to match the damage found.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected after a fall.The user has been re-implanted.
|
| |
|