| Model Number |
0684-00-0568-01U |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by customer that the sensation plus 7.5fr 40cc intra-aortic balloon (iab) had a fo sensor failure and all attempts at troubleshooting are unsuccessful.There is no harm to the patient due to the reported issue.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided, we will send a supplemental report with our additional findings.Due to characters limitations, e1 (initial reporter) - (b)(6).Complaint id - (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Updated fields - b4, g3, g6, h2, h6 (type of investigation, investigation conclusions), h11.Corrected field: d9.No decon or visual inspection performed since product was not returned and could not be evaluated.Therefore, we were unable to confirm or determine a root cause for the reported failure.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).Based on the rational provided above, no escalation to the capa process is required.Complaint id# (b)(4).
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Search Alerts/Recalls
|