MAUDE Adverse Event Report: PERFUSION SYSTEMS ACT PLUS INSTRUMENT; TIMER, CLOT, AUTOMATED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PERFUSION SYSTEMS ACT PLUS INSTRUMENT; TIMER, CLOT, AUTOMATED Back to Search Results
Model Number ACT200
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/29/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluation summary: the reported erroneous results between screen 1 and 2 was verified during service.Service technician observed no failure observed with the acttrac but after dismantling the device service technician found that the height of the lifting bar had to be adjusted and balanced on each track.The issue was resolved by disassembly of the device, cleaning the bar to remove any presence of blood and adjusting the lifting bar height.Post repair testing was performed per specifications.The instrument was serviced/analyzed in the facility by a field service technician.The instrument did not return to a medtronic facility for service/analysis.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Medtronic received information that during use of an act plus instrument, it was reported that the results were erroneous between screen 1 and 2.The use of the instrument was unknown.There was no adverse patient effect associated with this event.
 
Additional Manufacturer Narrative
Medtronic received additional information that there was no error code associated with this issue observed.The test results were not obtained.Correction b3: date of event updated.Correction g3: date mfr rec in initial report was confirmed to be 2025-01-29.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACT PLUS INSTRUMENT
Common Device Name
TIMER, CLOT, AUTOMATED
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key21410811
Report Number2184009-2025-00278
Device Sequence Number9455798
Product Code GKN
UDI-Device Identifier00643169178410
UDI-Public00643169178410
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K940426
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberACT200
Device Catalogue NumberACT200
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/27/2025
Supplement Date Received by Manufacturer05/12/2025
Initial Report FDA Received Date02/19/2025
Supplement Report FDA Received Date05/29/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/06/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-