| Model Number |
N/A |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/27/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and it will be evaluated.A supplemental will be submitted with evaluation results.
|
| |
|
Event or Problem Description
|
|
It was reported by customer that the picc fracture/leaking.Additional information received 02/07/2025: it was reported patients' picc had to be removed and another picc was inserted.Additional information received on 02/13/2025: it was reported the leak was external to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a leak was confirmed and was determined to be use related.The product returned for evaluation was one 4 fr dual lumen powerpicc sv catheter.The investigation findings are consistent with damage accumulated through flexural fatigue.Flexural fatigue may occur due to cyclic kinking of the catheter tube in which physiological, placement, usage, and mechanical factors may gradually form a crack(s) in the catheter.The returned product sample was evaluated and a split was observed in the catheter tubing just distal to the molded joint.The catheter split contained physical features associated with material fatigue, and the characteristics observed which supported this type of failure included: damage which was circumferentially aligned.Fracture edges which were rounded and polished due to repeated material wear.'c' shaped impressions leading into the fracture site which are consistent with material buckling due to movement which caused the fracture edges to be pressed together.Overall elliptical shape to the fracture cross-section (a result of repeated kinking of the tubing).An examination of the catheter structure revealed no potential damage/defect related to manufacture of the product.The damage location suggested that catheter securement, access and maintenance techniques may have contributed.This complaint will be recorded for future trending and monitoring purposes.
|
| |
|
Search Alerts/Recalls
|