Model Number G32224
Medical Device Problem Code Unsealed Device Packaging (1444)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/05/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Summary of event: as reported, prior to patient contact, a flexor balkin guiding sheath's package was not closed properly.Per the reporter, the sterile seal was missing on one side.The package was not ripped or torn.The device did not make patient contact.The patient did not require any additional procedures or experience any adverse effects due to this event.Investigation evaluation: reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturing instructions, and quality control procedures were conducted during the investigation.A visual inspection of the complaint device was also conducted.The unused complaint device was returned to cook for investigation.One seal was missing at the bottom of the pouch.Frosting was not noted where the seal should be.A document-based investigation evaluation was performed.A review of the device history record found no relevant non-conformance's on the lot.A review of complaint history found no additional complaints for this lot number.A search of all lots packaged and inspected by the same personnel as the complaint lot found no relevant non-conformance's or additional complaints.The product ifu states ¿supplied sterilized by ethylene oxide gas in peel-open packages.Intended for one-time use.Sterile if package is unopened and undamaged.Do not use the product if there is doubt as to whether the product is sterile.Store in a dark, dry, cool place.Avoid extended exposure to light.Upon removal from package, inspect the product to ensure no damage has occurred.¿ a review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The information provided upon review of the dmr, dhr, ifu, and investigation of the returned device suggests that there is evidence the device was manufactured out of specification; however, cook has concluded that this is an isolated incident.There is no evidence that additional non-conforming product exists in-house or in the field.Based on the information provided and the results of the investigation, cook has concluded that a manufacturing and quality control deficiency contributed to this event.Responsible personnel were notified.The risk analysis for this failure mode was reviewed and no additional escalation was required.The appropriate personnel have been notified, and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Additional Manufacturer Narrative
E1: customer country: (b)(6).H3: device evaluation anticipated, but not yet begun.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
As reported, prior to patient contact, a flexor balkin guiding sheath's package was not closed properly.Per the reporter, the sterile seal was missing on one side.The package was not ripped or torn.The device did not make patient contact.The patient did not require any additional procedures or experience any adverse effects due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXOR BALKIN GUIDING SHEATH
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key21411921
Report Number1820334-2025-00144
Device Sequence Number8215108
Product Code DYB
UDI-Device Identifier00827002322240
UDI-Public(01)00827002322240(17)270916(10)16250137
Combination Product (Y/N)N
PMA/510(K) Number
K142829
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG32224
Device Catalogue NumberKCFW-6.0-38-40-RB-BLKN-HC
Device Lot Number16250137
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/05/2025
Supplement Date Received by Manufacturer05/07/2025
Initial Report FDA Received Date02/19/2025
Supplement Report FDA Received Date05/13/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/16/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-