• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS ULTRASOUND, INC EPIQ ELITE; SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS ULTRASOUND, INC EPIQ ELITE; SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC Back to Search Results
Model Number EPIQ ELITE ULTRASOUND SYSTEM
Medical Device Problem Code Poor Quality Image (1408)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/22/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported the epiq elite ultrasound system along with an el18-4 transducer were producing images during a breast scan that were not able to diagnose anechoic cysts accurately.Patients had unnecessary aspirations of cysts.The customer is currently using an l12-5 transducer and obtaining additional images to prevent treatment of anechoic cysts.No further information was provided.Philips has started an investigation, and upon completion, a follow up report will be submitted.
 
Additional Manufacturer Narrative
A thorough investigation was performed to determine the cause of the reported problem.The investigation findings determined the system was working as designed with the autoscan feature.The customer was provided with recommendations to help modify settings to better suit their imaging preferences, and no further similar issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIQ ELITE
Common Device Name
SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC
Manufacturer (Section D)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
roxanne ramirez
22100 bothell everett highway
bothell, WA 98021
MDR Report Key21412239
Report Number3019216-2025-000041
Device Sequence Number13141304
Product Code IYN
UDI-Device Identifier00884838099111
UDI-Public(01)00884838099111
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K200304
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEPIQ ELITE ULTRASOUND SYSTEM
Device Catalogue Number795234
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/23/2025
Supplement Date Received by Manufacturer05/11/2026
Initial Report FDA Received Date02/19/2025
Supplement Report FDA Received Date05/18/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/05/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-